Specifications

A64192AA2 FINAL
2-3
Outlet Unit Error Recovery Codes
Outlet Unit Error Recovery Codes
2
1-19 SN19 Proximity sensor did
not detect a sample
tube carrier entering
the Outlet waiting
queue.
Outlet Unit C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-20 SN20 Outlet Transfer Arm
error finding the Y-axis
home position.
Outlet Unit C-5 1. Ensure the Outlet Transfer Arm movement is
not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
1-21 SN21 Outlet Transfer Arm
error finding the Y-axis
home position.
Outlet Unit C-5 1. Ensure the Outlet Transfer Arm movement is
not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
2-01 SL01 Outlet Transfer Arm Z-
axis cylinder SL01 did
not fully extend /
retract and misses
activating the limit
sensor.
Outlet Unit Transfer
Arm
1. Clear any obstruction.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
2-03 SL03 Stopper error at the
Outlet sample tube
carrier loading/
unloading position.
Outlet Unit B-3
b
1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-04 SL04 Stopper error at the
Outlet sample tube
carrier loading/
unloading position.
Outlet Unit B-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-05 SL05 Stopper error before
the Outlet sample tube
carrier loading/
unloading position.
Outlet Unit B-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-06 SL06 Stopper error at Outlet
waiting queue lane
Outlet Unit C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
Tab le 2. 1
Outlet Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution