Specifications
A64192AA2 FINAL
2-2
Outlet Unit Error Recovery Codes
Outlet Unit Error Recovery Codes
1-06 SN06 Outlet rack #2 from the
left not positioned
correctly.
Outlet Unit D-4/5 1. Ensure that Outlet rack #2 is in place and
correctly seated.
2. Press PAUSE/RUN to clear the error.
1-07 SN07 Outlet rack #3 from the
left not positioned
correctly.
Outlet Unit D-5 1. Ensure that Outlet rack #3 is in place and
correctly seated.
2. Press PAUSE/RUN to clear the error.
1-08 SN08 Outlet rack #4 from the
left not positioned
correctly.
Outlet Unit D-5 1. Ensure that Outlet rack #4 is in place and
correctly seated.
2. Press PAUSE/RUN to clear the error.
1-09 SN09 Sample tube carrier
error at Outlet bar-code
reader.
Outlet Unit C-3 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-10 SN10 Fiber optic sensor did
not detect a sample
tube at Outlet bar-code
reader BR01.
Outlet Unit C-3 1. Locate missing tube and place into the
waiting carrier.
2. Press PAUSE/RUN to clear the error.
1-11 SN11 Proximity sensor did
not detect a sample
tube carrier released
from the Outlet bar-
code reader.
Outlet Unit B2 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-12 SN12 Sample tube carrier
proximity sensor error
before the Outlet bar-
code reader.
Outlet Unit C-4 This error occurs during Power On, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
1-13 SN13 Sample tube carrier
proximity sensor error
at the Outlet waiting
queue.
Outlet Unit C-4 This error occurs during Power On, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
1-16 SN16 Sample tube carrier
error at the Outlet
return track.
Outlet Unit B-5 This error occurs during Power On, if the sensor
is defective.
1. Press PAUSE/RUN to clear the error.
1-18 SN18 Proximity sensor did
not detect a sample
tube carrier from the
return track.
Outlet Unit C-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently
twist and push the sample tube carrier in the
direction of least resistance.
3. Press PAUSE/RUN
to clear the error.
Tab le 2. 1 Outlet Unit Error Codes (Continued)
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution