Specifications

A64192AA2 FINAL
2-1
CHAPTER 2
Outlet Unit Error Recovery Codes
Outlet Unit Error Recovery Codes
The Outlet Unit Error Code Table 2.1 describes error messages generated by the LH 1500
Hematology Automation System. The table provides the system error codes, the related sensor, a
location reference point on the schematic diagram and a brief explanation of the possible problem
and possible solutions.
Tab le 2. 1 Outlet Unit Error Codes
Keypad
Display Sensor
a
Problem
Sensor
Diagram
Grid
Ref Solution
0-00 BZ00 Bar-code label did not
read or was incorrectly
positioned on tube.
Outlet Unit B-3 1. Ensure the bar-code is present and properly
positioned.
2. Turn the sample tube in the carrier at BR01
with bar-code facing the bar-code reader.
3. Press PAUSE/RUN to clear the error.
0-01 BZ01 Bar-code label did not
read or was incorrectly
positioned on tube.
Outlet Unit B-3 1. Ensure the bar-code is present and properly
positioned.
2. Turn the sample tube in the carrier at BR01
with bar-code facing the bar-code reader.
3. Press PAUSE/RUN to clear the error.
1-01 SN01 Outlet Transfer Arm
error finding the X-axis
home position.
Outlet Unit D-3 1. Ensure there are no obstructions in the path
of the Outlet Transfer Arm.
2. Press PAUSE/RUN to clear the error.
If the problem persists, call your Beckman
Coulter Representative.
1-02 SN02 Outlet Transfer Arm
error finding the X-axis
home position.
Outlet Unit D-3 1. Ensure there are no obstructions in the path
of the Outlet Transfer Arm.
2. Press PAUSE/RUN to clear the error.
3. If the problem persists, call your Beckman
Coulter Representative.
1-05 SN05 Outlet rack #1 from the
left not positioned
correctly.
Outlet Unit D-3 1. Ensure that Outlet rack #1 is in place and
correctly seated.
2. Press PAUSE/RUN to clear the error.