Error Recovery Procedures
Secondary Analysis Mode on the LH 1500 System
1
Figure 1.2 LH Cassette Handling Mechanism
Secondary Analysis Mode on the LH 1500 System
The use of Secondary Mode allows analysis of specimens without track processing. Secondary Mode
is the quickest mode of analysis for STAT specimens when numerous specimens are being processed
by the LH 1500 System.
1
Press STOP one time ONLY on the LH 750 Analyzer's keypad.
2
When "READY" appears on the LH 750 Analyzer use the bar-code wand and scan the sample ID
in the Bar-code field into the LH 750 Workstation Command Center, or input the sample ID into
the LH 750 Analyzer numeric keypad.
3
Aspirate the sample using the secondary aspiration probe.
4
Repeat steps 2 - 3 until all samples are analyzed.
NOTE
If numerous samples are to be analyzed in Secondary Mode, proceed to DISABELING THE
LH 1500 CONNECTION UNIT Procedure. Disabling the Connection Unit will allow the workload to be
diverted to the other instrument. Processing samples while PAUSED will back-up the workload during
Secondary Analysis.
5
After the cycle completes and results are received at the LH 750 Workstation ("READY" appears
on the LH 750 Analyzer), press
START/CONT on the LH 750 Analyzer's keypad to continue
automatic mode processing.