Specifications

A64192AA2 FINAL
1-10
Error Recovery Procedures
Loading Cassettes in the Connection Unit
Loading Cassettes in the Connection Unit
Proper loading of the cassettes minimizes system errors and allows the Connection Unit to
accurately organize the cassettes.
1
Before loading cassettes into a LH Connection Unit, ensure that the:
LH 1500 robotic line has initialized with no errors.
LH connection keypad LED indicates the AUTO status.
LH connection Unit has not detected an error (if possible).
2
Remove the right-side tabletop access cover of the Connection Unit and set aside. The
Connection Unit will Pause.
3
Load the first cassette at position #5 (the label must face away from operator). Refer to
Figure 1.2, LH Cassette Handling Mechanism.
4
Replace the right-side tabletop access cover back into the proper safety position; make sure the
sensor in the upper left corner is properly positioned.
5
Press the PAUSE/RUN button on Connection Unit. Allow the cassette to advance to position
number six.
6
Repeat steps 2 through 5 swiftly until fourth (last) cassette is loaded onto Connection Unit.
Timing is crucial. Do not exceed the maximum number of four cassettes.
NOTE
Do not place more than one cassette at a time on the Connection Unit.
Do not insert next cassette if another cassette is in position number four, wait until cassette
advances to position number six.
7
Ensure the cassettes have fully cycled and no errors are displayed on the keypad.
NOTE
Home cassette positions are indicated by red outlined areas (Position number one, Position
number five, Position number six and Position number eight).