Specifications
A64192AA2 FINAL
1-9
Error Recovery Procedures
Removing Cassettes from the Connection Unit
1
6
Remove third cassette at position number one. At the same time remove the fourth (last)
cassette at position number six. Proceed to Step 9.
7
Remove the right-side tabletop access cover and remove the front covers of the Connection
Unit, and set aside.
8
Remove all four of the cassettes from the Connection Unit.
NOTE
Once a specimen is scanned by the bar-code reader inside the Connection Unit, the Line
Controller assumes analysis is complete.
• During Error Recovery there may be unanalyzed specimens in a Connection Unit or on a LH 750.
Unanalyzed specimens must be analyzed by Secondary Analysis Mode or STAT-CASS Mode. If
not, the unanalyzed specimens will go to the Stockyard or Outlet if placed back on the LH 1500
System.
• During Error Recovery specimens that have been analyzed by the LH 750, but not routed out of
the Connection Unit, can be placed back on the LH 1500 System, and will be properly routed to
the Stockyard or Outlet.
9
Replace the front covers back into their proper positions and/or the right-side tabletop access
cover.
10
Press the PAUSE/RUN button on Connection Unit. Ensure there are no errors displayed on the
keypad.
NOTE
If errors persist due to a cassette jam, see Loading Cassettes in the Connection Unit. Multiple
Errors and Alarms may occur. Continue to PAUSE/RUN and/or load cassettes until errors are cleared.
Figure 1.1 Cassette Handling Mechanism