Specifications

A64192AA2 FINAL
1-6
Error Recovery Procedures
Error Recovery Startup
9
Press the red power OFF button on the Inlet Unit and turn off the automation line.
NOTE
Wait 60 seconds before pressing the green power ON button.
Error Recovery Startup
The Error Recovery Startup procedure brings the LH 1500 system online; powering on the
automation line, re-initializing robotics, restoring LIS interfacing and command to the Line
Controller.
1
Complete System Shutdown Procedure.
2
Ensure the LH 755 Analyzers are powered up including: LH 750 Analyzers, Workstations,
Slide Makers and Slide Stainers, if applicable.
3
On the LH Unit Blocks and Tracks remove any tubes and/or foreign objects.
4
Ensure the yellow emergency Stop buttons are in the "Out" position (not pushed) on all LH 1500
system components.
5
Press the green power ON button on the Inlet Unit to power on the automation line.
6
Select System Operation > Startup on the Main Menu Screen Menu Bar. The Startup screen
displays and shows the options available for managing sample data while performing Startup
on the LH 1500 System.
7
The following options are available:
Delete Processed Samples Only - (default selection) This function looks at all processed
samples, and unprocessed samples more than 72 hours old, and deletes them (includes
exchanged racks). This deletion does not affect the stockyard or UDR racks, although
samples in the Error rack and Pending rack will be deleted.
Do Not Delete - saves all of the data currently in the Line Controller.
Delete All Sample Data - deletes all data currently in the Line Controller. Select
<Do Not
Delete>
8
Select <OK> in the Startup screen to have the robotics initializes.