Technical data

Settings MIQ/CR3; DIQ/CR3
4 - 34
ba76032e01 01/2012
Application example Feedback control of the D. O. concentration
Sensor: TriOxmatic 700 IQ (measuring range: 0 ... 60 mg/l)
The control parameters form the following (negative) characteristic
curve:
Fig. 4-19 Characteristic curve for application example
The controller works with the following gain:
Control parameters Value
Nominal value 4mg/l
Xp 10 % of the measuring range or 6 mg/l
Imin 8mA
Imax 14 mA
Io 12 mA
ti 0 s (no I controller part)
td 0 s (no D controller part)
0
8
14
12
Imax
Imin
Current I [mA]
Measured value [mg/l]
428
Io
Nominal value
Xp
mg/l
mA
1
mg/l6
mA6
K ==