Technical data

-23-
Special Status and relevant Error Messages
The table below describes those special status which cause the yellow fault LED to be lit
permanently. For a better diagnosis of a transmitter without display you should either read the error
messages below from the remote control RC2 or analyse the values from the current output.
No
Display
green
LED
yellow
LED
Output
Cause Action
01
"
TEST
"
On On 0 mA Memory test when booting the system finishes automatically after 6
seconds
02
"
Load
"
Operational
para.
"
AdJ
"
flashing On 1,6 mA Booting the system at start or after
change of gas (see page 8)
turns to sensor warm-up
automatically
03
Countdown
of seconds
flashing On 1,6 mA Sensor warm-up period finishes automatically after 120
seconds.
10
"
TEST
"
flashing
Off
glows 0 mA Too low supply voltage increase supply voltage
11
"
CHNLM
"
"
SENS
"
flashing
shortly
(single
flash)
On 1,2 mA Sensor lifetime expired Replace sensor
12
"
CHEC
"
"
L MAS
"
Twin
flash
On 1,2 mA After sensor replacement:
Sensor is not specified for the gas
Replace sensor again or
select different gas; see page 19
„Sensor Replacement“
13
"
CHEC
"
"
SCAL
"
Twin
flash
On 1,2 mA After sensor replacement:
Detection range setting of sensor and
transmitter do not match for the gas
Check and, if necessary, change
detection range SCAL
See page
19 „Sensor Replacement
14
"
SYS
"
"
ERR.1
"
Off On 1,2 mA Fault during RAM access
Re-start transmitter. If error is
reported again, replace transmitter
15
"
SYS
"
"
ERR.2
"
Off On 1,2 mA Fault during ROM access
Re-start transmitter. If error is
reported again, replace transmitter
16
"
SYS
"
"
ERR.3
"
Off On 1,2 mA Fault during EEPROM access (internal)
Re-start transmitter. If error is
reported again, replace transmitter
17
"
SYS
"
"
ERR.4
"
Off Off 1,2 mA Reserved
Re-start transmitter. If error is
reported again, replace transmitter
18
"
SYS
"
"
ERR.5
"
"
SYS
"
"
ERR.5
"
On 1,2 mA Fault during regulation of sensor voltage
19
Off
Off On 1,2 mA Fault during regulation of hardware
zeropoint
20
"
SYS
"
"
ERR.7
"
Off On 1,2 mA Temperature measurement is not
plausible
Re-start transmitter. If error is
reported again, replace transmitter
21
"
SENS
"
"
ERR.1
"
"
SENS
"
"
ERR.1
"
On 1,2 mA Sensor is missing
Fit sensor (automatic re-start)
22
"
SENS
"
"
ERR.2
"
Off On 1,2 mA Fault during EEPROM access (sensor)
Replace sensor (automatic re-
start)
23
"
SENS
"
"
ERR.3
"
Off On 1,2 mA Wrong sensor/parameter memory
24
"
ADv
"
"
ERR.1
"
Off On 1,2 mA Error A/D converter (sensor voltage) Re-start transmitter.
If error is reported again, replace
transmitter.
25
"
ADv
"
"
ERR.2
"
"
ADv
"
"
ERR.2
"
On 1,2 mA Error A/D converter (NTC) Re-start transmitter.
If error is reported again, replace
transmitter.
26
"
ADv
"
"
ERR.3
"
Off
On 1,2 mA Error A/D converter (sensor current) Re-start transmitter.
If error is reported again, replace
transmitter.