User`s manual

93
This message alerts you to the Air-In-Line Sensor being set to “OFF.” The message will
require answering “yes” or “no” to whether the administration set being used has an air-
eliminating filter. If you press the “YES” key to indicate that there is an air-eliminating filter in
use, the Air-In-Line sensor will be completely disabled and the pump will start the infusion.
If you press the “NO” key to indicate a filter is not in use, the Air-In-Line sensor will be set to
detect the presence of any contiguous 2-milliliter air in the line. The alert message below will
then briefly appear on the screen. If the alert occurred at the Run Options screen, the pump
will start the infusion.
Air-In-Line Sensor Set at 2.0ml Alert Screen
Also, when the Air-In-Line sensor is completely disabled and the infusion goes to an
“Infusion Complete” or “Empty Bag” status, the Air Sensor setting will automatically change
to 2.0 ml.
6 NXT DOSE (Next Dose): Allows the time of the next dose to be changed. The current
time-remaining-until-the-next-dose is displayed. Use the numeric keys to change the time for
the start of the next dose. If correct, press the “YES/ENTER” key to accept, and move to
the next field. The range for this field is 0:00 to 24:00 HH:MM.
Note: If there is a dose currently in progress and the NXT DOSE feature is used, the current
dose will be terminated and the cycle will continue at the KVO rate until the time for the
new next dose.
7 AUDIO (Audio Volume): This feature allows you to adjust the audio volume of the pump’s
alarm setting. You can enter a number from 9 (the loudest setting) to 1 (the quietest setting).
When you select a number, the pump will beep at that audio level. Press the
“NO/CHANGE” key and enter a new setting (1 - 9). Repeat until the desired setting is
reached, and then press “YES/ENTER” to select it and move the cursor to the next field.
8 PWR CK? (Power Check): This feature allows the user to recheck the battery charge. The
display gives the active power source of the pump, whether internal batteries or external
power. If the active source is the batteries, a bar graph will indicate approximately how much
charge remains in the batteries. To access this information, use the “NO/CHANGE” key to
change from the default setting of “OFF” to “ON,” and press “YES/ENTER”. The Power
Source Screen will display for five seconds and return to the Options Menu.
9 ACCEPT?: If the changes made in the Options Menu are acceptable, press the
“YES/ENTER” key. If the changes are not acceptable, use the arrow keys to cursor up to
any field for corrections, or abort all changes made in this menu by pressing the “NO” key
and return to the previous screen.
There are two remaining selections to consider from the “Run Options Screen”: NO to
REVIEW and RUN to START.
Intermittent Therapy