Note: You may elect to skip the dose rate and enter the dose time first to let the pump
automatically calculate the dose rate.
4 DOSE Time: Using the numeric keys, enter the amount of time over which each dose is to be
infused. The range for this field is 0:01 to 99:59 HH:MM. When correct, use the
“YES/ENTER” key to accept this field, and to move on to the next.
Note: If you have entered the dose amount and rate, the pump will automatically calculate the
dose time.
5 DOSE Freq (Dose Frequency): Represents the time from the start of one dose to the start of
the next (e.g., every six hours). Use the numeric keys to enter the dose frequency, and when
correct, press the “YES/ENTER” key to accept the cycle time and to move on to the next
field. The range for this field is 0:01 to 99:59 HH:MM.
6 KVO Rate (Keep Vein Open Rate): This field requires data entry using the numeric keys to
set the rate at which fluid will be delivered in a “Keep Open” state. The KVO Rate will be
used between doses. The pump can be set from 0 to 10 ml/hour, but cannot be set at a rate
to exceed the therapy-operating rate. When the desired KVO rate has been entered, press the
“YES/ENTER” key to accept and move the cursor to the next field.
When a KVO rate is programmed for the Intermittent Therapy, the
patient is not disconnected between dosing. The pump
automatically cycles through dosing and KVO. A single beep alerts
the user to the change between dosing and KVO. If the KVO rate is
set at “0”, the pump can be off between doses. The internal time
clock of the pump continues to monitor the dosing cycle.
Note: The next three fields are automatically calculated based on the information already
programmed and are helpful in providing additional information for managing an
intermittent therapy. At each of these fields, when the information is correct, press the
“YES/ENTER” key to accept and move to the next field.
7 DOSE/BAG: Displays the number of doses that can be delivered from the BAG Volume
given. You can change this number to a lower number of doses per bag but cannot change to
a greater number. When correct, press the “YES/ENTER” key to accept, and move to the
next field. The range for this field is 1 to 200.
8 REQ Vol (Required Volume): The volume needed for the programmed number of doses,
including the amount needed for KVO rates (if a KVO rate is programmed). The pump
calculates this volume for you based on previously programmed information, and you cannot
change the data in this field. Use the down arrow or the “YES/ENTER” key to move to the
next field. The range for this field is 1 to 9999.
9 TOT Time (Total time): Time required to deliver the number of doses programmed. The
pump calculates this time based on previously programmed information and you cannot
change data in this field. Use the down arrow or the “YES/ENTER” key to move to the
next field. The range for this field is 0:01 to 288:00 HH:MM.
Note: A DOSE is always followed by a KVO period. Therefore, after the last dose, the KVO
will run until the dose frequency time has expired.
IMPORTANT
INFORMATION