Intermittent Therapy
Intermittent Therapy is designed to deliver programmed intervals and rates of specified amounts of infusates and to
optionally deliver small amounts of the infusate between doses to keep the access site patent.
egin programming an Intermittent Therapy by selecting “Intermittent” from the “Select
Therapy Screen”. The following display screen will appear, requiring you to enter
information needed to program this therapy.
Prescription Menus
Intermittent Pre-Prescription Menu Screen
These menu items allow you to customize the program to match each individual prescription. The
pump’s default settings are displayed in the left screen. Each field is individually explained below.
Note: Some of the lines are abbreviated to fit on the display screen; the full descriptions appear
below.
1 UNITS: Allows you to select from units of ml, mg, or mcg, to match the prescription. The
pump will allow a volume range of 0.1 to 400 ml/hr; therefore, if mg or mcg is selected, the
concentration requested in the next field will determine the upper and lower programming
limits for mg and mcg. Use the “NO/CHANGE” key until the desired “units” appears on
the display, and then press the “YES/ENTER” key to accept and move to the next field.
2 CONCEN (Concentration): This field is active whenever mg or mcg is selected in the “units”
field and requires data entry of the concentration of mg or mcg per ml. When the correct
data is entered, press “YES/ENTER” to accept, and move the cursor to the next field. If ml
is chosen in the “units” field, this line will not display. The range for this field is 0.1 to 999.
3 DELAY: This field allows you to delay the start of the infusion from one minute up to
twenty-four hours. If the infusion is NOT to be delayed, accept the default setting of “OFF”
by pressing the “YES/ENTER” key, and the cursor will move to the next field. If the
infusion IS to be delayed until a later time, press the “NO/CHANGE” key. The delay start
will change to “ON.” Now, when you press the “YES/ENTER” to accept, the following
Chapter
7
B