User`s manual

83
1 “INFUSION COMPLETE.”
2 A graphic representation (completely filled in bar graph) indicating infusion complete, or the
smaller “dot” configuration (shown here) if a KVO RATE is programmed.
3 TOT Vol (Total Volume): Displays the total volume infused.
4 KVO Rate: The preset KVO rate displays.
5 Infusion Complete is considered an alarm status, and the audio alarm will beep periodically
(every 3 seconds if there is no KVO, every 20 minutes if there is a KVO) to alert the user to
this condition. Additionally, the LED lights blink to alert you to the status of the pump. The
green light blinks when the KVO rate is infusing (if one is programmed) and the red light
blinks to indicate that the “Infusion Complete” alarm exists.
Stopping the Infusion
To halt the infusion at “Infusion Complete”, press the “PAUSE” key and then choose to either:
1 Turn off the pump by pressing the “OFF” key, or
2 Select one of the options from the “Pause” screen:
Pause Menu Screen
a. REPEAT Rx: To repeat the present therapy and add another IV container with the same
prescription, press the “YES/ENTER” key at this field. If more than one therapy is
programmed, a “Select Therapy Screen” will list all preprogrammed therapies. Select
which therapy to repeat by moving the cursor to the therapy of choice and pressing the
“YES/ENTER” key.
The pump will then display the program screens with the last programmed values of the
therapy and the cursor will automatically advance through the prescription menus. You
can stop the auto-advance function by pressing any key (except ON/OFF) and then
complete the confirmation of the prescription manually by pressing the “YES” key at the
remaining fields.
b. NEW PROGRAM: To program an entirely new prescription or different therapy, press
the “YES/ENTER” key at this field and the display will return to the “Select Therapy
Screen.” (The lock setting can control access to this function.)
Note: The therapy is complete; therefore, RESUME is not an option at this time and is not
displayed.
TPN Therapy