User`s manual

80
1 The first field displays the “RATE” of the infusion. If the therapy is in a ramp up or down
phase, the rate will change accordingly to reflect the gradual increase or decrease in rate. The
Rate toggles with the Volume to Be Infused.
2 The second field is a graphic bar representing the percentage of the completed infusion. It
also depicts the up or down ramping phase with gradually taller or shorter lines on the bar
graph. In CMS-enabled pumps, the graphic bar will toggle with Line Pressure (“L
Press”) which displays the current patient side (downstream) pressure in the
administration set in millimeter of mercury (mmHg) units.
3 The third field displays total volume infused.
4 The fourth field displays the remaining time to the completion of the infusion. This toggles
with the remaining volume.
5 In addition to the display screen accumulating information regarding the status of the
infusion, the green LED light blinks whenever the pump is infusing.
The information on the run screen will alternate between a four-line display and a large-font
display of each individual line of information on the screen.
TPN Information Screens
Interrupting an Infusion
The infusion can be paused at any time by pressing the “PAUSE” key. The yellow “Standby”
LED light blinks, and the following screen displays:
Pause Menu Screen