User`s manual
79
Also, when the Air-In-Line sensor is completely disabled and the infusion goes to an
“Infusion Complete” or “Empty Bag” status, the Air Sensor setting will automatically change
to 2.0 ml.
7 AUDIO (Audio Volume): This feature allows you to adjust the audio volume of the pump’s
alarm setting. You can enter a number from 9 (the loudest setting) to 1 (the quietest setting).
When you select a number, the pump will beep at that audio level. Press the “NO” key and
enter a new setting (1 – 9). Repeat until the desired setting is reached, and then press
“YES/ENTER” to select it and move the cursor to the next field.
8 PWR CK? (Power Check): This feature allows you to periodically recheck the battery charge.
It will display the active power source of the pump, whether internal batteries or external
power. If the active source is the batteries, a graph will indicate approximately how much
charge remains in the batteries. To access this information, select “ON,” and press
“YES/ENTER”. The Power Source Screen will display for five seconds and return to the
Options Menu.
9 ACCEPT OPT?: If the changes made in the Options Menu are acceptable, press the
“YES/ENTER” key. If the changes are not acceptable, use arrow keys to cursor up to any
field for corrections, or abort all changes made in this menu by pressing the “NO” key and
return to the previous screen.
There are two remaining selections to consider from the “Run Options Screen”: NO to
REVIEW and RUN to START.
Reviewing or Changing the Program
To review or make any changes in the prescription programming, press the “NO/CHANGE”
key at the Run Options screen, and the prescription screens will reappear (Changes may be
inhibited by lock setting.) Changes to the prescription can be made before the start of the
therapy; however, once a therapy has started, no changes can be made to the Pre-Prescription
Screen. It will appear only for review purposes. Attempts to change the Pre-Prescription Menu
during a new program review will cause an alert screen to be displayed, notifying the programmer
that changes will require reprogramming of the prescription.
Starting the Infusion
RUN: When all parameters are correct and the pump has been primed properly and connected
to the access site, press “RUN” at the Run Options screen to start the therapy. If a delayed start is
programmed, the “Delay Start Count Down Screen” will appear. If no delay start is programmed
or the delay has elapsed, the pump will begin infusing, and the following screen displays the status
of the therapy:
TPN Run Screen
This screen will continue to display as long as the infusion is running. Notice the information on
each field:
TPN Therapy