User`s manual
77
pump has been set into a locked condition, any subsequent request to change the lock will
require a clinician access code. (See “Clinical Information and Access Code” chapter for
access code information, lock setting table and available features for each lock setting).
2 HRLY TOT (Hourly Totals): To view hourly infusion totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. The following screen will appear:
Hourly Totals Screen
a. The screen displays hourly infusion delivery for each hour (00-23) of the last 24
hours of therapy in the unit selected.
a. The highlighted field has a marker ◄
indicating the current hour in progress.
b. Use the up arrow key to scroll back through the last 24 hours.
c. If the infusion was not running, a series of ------ will appear for that hour.
d. To exit this screen, press the NO/CHANGE key.
3 CLR HRLY (Clear Hourly Totals): To clear Hourly Totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. The following screen will appear:
Clear Hourly Totals Screen
a. To clear the Hourly Totals, press the “YES/ENTER” key. All Hourly Totals will be
cleared and the cursor will return to the Options menu.
b. To maintain the Hourly Totals, press the “NO/CHANGE” key. Hourly Totals will be
maintained and the cursor will return to the Options menu.
c. A lock level of 2 or 3 will require an access code entry before allowing “Hourly Totals”
reset.
4 DN Ramp (Early Down Ramp): During a TPN infusion, it may be necessary or desirable to
stop the level rate of infusion and begin an early down ramp process. The range for this field
is 0:00 to 9:00 HH:MM. Selecting early ramp down by selecting “ON” and pressing the
“YES/ENTER” key when the highlight bar is on this field brings a second screen as seen
below:
TPN Therapy