User`s manual

76
7 To continue priming, press the PRIME key again. If priming is complete, press
“YES/ENTER” to exit and return to the Run Option screen. The total amount of solution
used in this priming session displays on the bottom field.
Note: The amount of fluid used in the priming process will be deducted from the Bag Volume
Total but will NOT be counted in the amount infused because this amount is not to be
delivered to the patient. The bag volume should still be equal to or greater than the
Amount TBI before you can proceed with the infusion. If it is not, an alert will appear
stating, “AMT TBI GREATER THAN BAG VOL”. You have the option to continue
by pressing the “YES/ENTER” key or return to the “Run Options” screen by pressing
the “NO” key.
Options
Once the prescription is programmed, there are more customizing features available for each
therapy. Press the “OPTIONS” key from the Run Options screen to access these options. The
OPTIONS menu can also be accessed while the pump is running.
Note. After the infusion has started, any changes made through the OPTIONS menu involving:
“HRLY TOT”, and
“CLR HRLY”
will return you directly to the “ACCEPT OPT?” field.
The following screen displays the TPN Therapy “Options” available. When you open this menu,
use the up and down arrows to move the highlight bar to the desired option. Lock settings may
require an access code before the option can be changed. Any new setting, except “CLEAR
HOURLY” screen acceptance becomes effective when the “Options” screen is exited by pressing
“YES/ENTER” at the “ACCEPT OPT?” field.
TPN Options Menu
Note: The pump’s default settings are indicated above in the left screen.
1 LOCK: When the lock is set to “OFF ” the clinician has access to all of the pump functions.
After the pump is completely programmed, change the lock to meet the security needs of this
therapy. There are four choices: OFF, 1, 2, or 3. Each level higher limits the available
options. To change the lock setting, press the “NO/CHANGE” key until the correct lock
setting appears. Then, press “YES/ENTER” key to accept the new lock selection. Once the