before they are connected to the access site. Priming may have been done by gravity when the
administration set was first connected to the bag (see Chapter 1, “Preparing Medication for
Infusion”). However, if it was not done then or if there are any remaining air bubbles to be
removed, the pump’s “PRIME” function will assist in this process. To prime the set by using the
pump, follow the directions below.
Note: The lock setting may restrict the ability to prime using the pump.
1 Press and release the “PRIME” key at the “Run Options” screen.
2 The following screen appears:
Prime Direction Screen
3 Disconnect the administration set from the access site using aseptic protocols
Priming with the set connected to the patient could result in overdose
and may cause injury or even death to the patient.
4 Release any clamps on the administration set.
5 Press and hold the “PRIME” key. Fluid will be pumped at a rate of 300 ml/hr through the
administration set as long as the PRIME key is pressed or until 6 ml is pumped. The pump
will then beep once and require repressing and holding the “PRIME” key again until 6 more
milliliters is delivered. Continue this process until all air has been purged from the IV bag and
the administration set.
Note: If the bag volume is depleted during priming, a subsequent priming event cannot be
initiated.
While the “PRIME” key is depressed, the following screen appears:
The amount used for the prime session is accumulated and displayed on the bottom field of
this screen.
6 When priming is complete, release the “PRIME” key. The following screen will display:
IMPORTANT
INFORMATION
TPN Therapy