User`s manual
71
TPN Therapy with Automatic
Ramping
This therapy is designed to allow a level rate of infusion of parenteral nutritional products with the option of tapering
at the beginning, end, or both of the infusion. It also has an early ramp-down feature.
egin programming a TPN Therapy by selecting “TPN” from the “Select Therapy Screen”.
The following display screen will appear, requiring you to enter information needed to
program this therapy.
Prescription Menus
TPN Pre-Prescription Menu Screen
This menu item allows you to customize the program to match each individual prescription. The
pump’s default setting is displayed in the left screen. Each field is individually explained below:
1 DELAY (Delay Start): This field allows you to delay the start of the infusion from one minute
up to twenty-four hours. If the infusion is NOT to be delayed, select “OFF” and press the
“YES/ENTER” to enter. The cursor will move to the next field. If the infusion IS to be
delayed until a later time, select “ON” and press “YES/ENTER” to accept. The following
screen will appear:
TPN Delay Start Setting
The time format (12 hr AM/PM or 24 hr Military) selected in the BIOMED SETUP Menu
Chapter
6
B
TPN Therapy