b. If the incomplete therapy is to be continued, press “NO” and the Pause Menu screen will
be displayed, then press “YES” and the Run Options screen will be displayed. To resume
the unfinished therapy, press “RUN”.
Entering a New Program
3 NEW PROGRAM: To program an entirely new prescription or different therapy, press the
“YES/ENTER” key at this field (the ability to change these parameters may be controlled by
the lock setting). The display will return to the “Select Therapy Screen” if more than one
therapy mode is enabled, or to the “Pre-Prescription Screen” of the single therapy mode. If
the current container has not completely infused, the following screen will appear:
a. This screen is a reminder that the infusion is not complete. If you want to return to the
current container and finish infusing it, press the “NO/CHANGE” key to return to the
Pause Menu screen.
b. To abort the current container and enter a new program, press the “YES/ENTER” key.
If a prescription has previously been entered for the therapy mode you wish to program,
the following screen will appear:
c. If you want to save the current prescription, press the “NO/CHANGE” key to return to
the Pause Menu screen. If you wish to erase the current prescription and enter a new
program, press the “YES/ENTER” key to proceed. All previous prescription values will
be erased.
Note: Selecting “NEW PROGRAM” will also reset the OPTIONS selected to their default
settings except for the LOCK and AUDIO settings which must be changed manually.
Infusion Complete
When the infusion is complete, the audio alarm will beep, and the following screen will display:
Note: To temporarily silence the alarm, press the “Silence” key.
PCA Infusion Complete
PCA Therapy