User`s manual
67
CMS-enabled pumps, the graphic bar will toggle with Line Pressure (“L Press”)
which displays the current patient side (downstream) pressure in the administration
set in millimeter of mercury (mmHg) units.
3 The third line displays the total volume “TOT Vol” infused and, if mg or mcg is the selected
unit, this field will toggle to display the total mg or mcg infused “TOT INF” as well.
4 The fourth line displays the number of boluses delivered and number of boluses attempted
per bag volume. If the patient attempts a bolus before the allowed amount of time set
between boluses, an alert message will display “Bolus Requested,” and only the second
number will increase. Also, toggling with number of boluses delivered/requested is the
remaining volume left in the bag.
5 In addition to the display screen, information on the status of the infusion is provided by the
three LED lights. The green LED light blinks whenever the pump is infusing. If the Basal
rate is set at “0”, the yellow LED light will blink along with the green light until the bolus is
actually being delivered. While the bolus is infusing, only the green light will blink.
The information on the run screen will alternate between a four-line display and a large font
display of each individual line of information on the screen.
PCA Information Screens
Interrupting an Infusion
The infusion can be paused at any time by pressing the “PAUSE” key. The yellow “Standby”
LED light blinks, and the following screen displays:
Pause Menu Screen
PCA Therapy