User`s manual
63
d. To exit the shift total feature without clearing the totals, press the “YES” key at the
highlighted EXIT field.
e. To clear the Shift Totals, press the “NO” key at the EXIT field. The field will change to
read “CLEAR”, press the “YES” key now to clear. A lock level 3 will require an access
code entry before allowing the user to clear the Shift Totals.
4 HRLY TOT (Hourly Totals): To view hourly infusion totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. Either of the following two screens
will appear:
Hourly Totals Screen
or
The screen displays hourly infusion delivery for each hour (00-23) of the last 24 hours of
therapy in the unit selected.
a. The highlighted field has a marker ◄ indicating the current hour in progress.
b. Volume or Amount will show in the first column, BG (Bolus Given) in the second
column and BA (Bolus Attempted) in the third column.
c. Use the up arrow key to scroll back through the last 24 hours.
d. If the infusion was not running, a series of ------ will appear for that hour.
e. To exit this screen, press the NO/CHANGE key.
5 CLR HRLY (Clear Hourly Totals): To clear Hourly Totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. The following screen will appear:
Clear Hourly Totals Screen
a. To clear the Hourly Totals, press the “YES/ENTER” key. All Hourly Totals will be
cleared and the cursor will return to the Options menu.
b. To maintain the Hourly Totals, press the “NO/CHANGE” key. Hourly Totals will be
maintained and the cursor will return to the Options menu.
PCA Therapy