User`s manual
61
Note. After the infusion has started, any changes made through the OPTIONS menu involving:
• “Clin DOSE”,
• “SHFT TOT”
• “HRLY TOT”, and
• “CLR HRLY”
will return you directly to the “ACCEPT OPT?” field.
The following screen displays the PCA Therapy “Options” available. When you open this menu,
use the up and down arrows to move the highlight bar to the desired option. Lock settings may
require an access code before the option can be changed. Any new setting, except “CLEAR
HOURLY” screen acceptance and “SHIFT TOTAL” clear screen acceptance becomes effective
when the “Options” screen is exited by pressing “YES/ENTER” at the “ACCEPT OPT?” field.
PCA Options Menu
Note: The pump’s default settings are indicated above in the left screen.
1
LOCK: When the lock is set to “OFF,” the clinician has access to all of the pump functions.
After the pump is completely programmed, change the lock to meet the security needs of this
therapy. There are four choices: OFF, 1, 2, or 3. Each level higher limits the available
options. To change the lock setting, press the “NO/CHANGE” key until the correct lock
setting appears. Then, press “YES/ENTER” key to accept the new lock selection. Once the
pump has been set into a locked condition, any subsequent request to change the lock will
require a clinician access code. (See “Clinical Information and Access Code” chapter for
access code information, lock setting table and available features for each lock setting).
2 ClinDOSE? (Clinician Dose): Allows an additional prescribed dose to be administered by a
clinician at any time except during a Load Dose or Bolus Dose. The Maximum Clinician
Dose is determined by the administration route (see Figure 5.1).
Clinician Dose Menu
PCA Therapy