PCA or Patient Controlled Analgesia is designed for therapies requiring a continuous rate of infusion, patient
controlled demand boluses, or both.
egin programming a PCA Therapy by selecting “PCA” from the “Select Therapy Screen”.
The following display screen will appear, requiring you to enter information needed to
program this therapy.
Prescription Menus
PCA Pre-Prescription Menu Screen
These menu items allow you to customize the program to match each individual prescription. The
pump’s default settings are listed in the left screen. Each field is individually explained below.
Note: Some of the lines are abbreviated to fit on the display screen; the full descriptions appear
below.
1 UNITS: Allows you to select from units of mg, mcg, or ml depending on how the
prescription is written. In the PCA therapy, the pump will allow rates as defined in figure 5.1;
therefore, if mg or mcg is selected, the concentration requested in the next field will determine
the upper and lower programming limits for mg and mcg. Use the “NO/CHANGE” key
until the desired “units” appears on the display, and then press the “YES/ENTER” key to
accept and move to the next field.
2 CONCEN (Concentration): This field is active whenever mg or mcg is selected in the
“units” field and requires data entry of the concentration of mg or mcg per ml. When the
correct data is entered, press “YES/ENTER” to accept, and move the cursor to the next
field. If ml is chosen in the “units” field, this line will not display. The range for this field is
0.1 to 999.