User`s manual

48
d. To exit this screen, press the NO/CHANGE key.
3 CLR HRLY (Clear Hourly Totals): To clear Hourly Totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. The following screen will appear:
Clear Hourly Totals Screen
a. To clear the Hourly Totals, press the “YES/ENTER” key. All Hourly Totals will be
cleared and the cursor will return to the Options Menu.
b. To maintain the Hourly Totals, press the “NO/CHANGE” key. Hourly Totals will be
maintained and the cursor will return to the Options Menu.
c. A lock level 2 or 3 will require an access code entry before allowing hourly total reset.
4 TitrRx (Titrate Prescription): If this feature was not
selected in the Pre-Prescription Menu, it
will not appear on the Options Menu. If a Titrate prescription was programmed and if the
therapy is to be titrated now, select “ON” and then press the “YES/ENTER” key to accept
and to go to the Continuous titration menu:
Continuous Titration Menu
This screen allows the user to change the rate within the predefined prescription limits.
a. RATE: Using the numeric keys, enter the new desired rate. When the new rate has been
entered, press the “YES/ENTER” key to move to the next field. If the titration option is
used to change the rate, the new rate will then appear on the run screen and on the
prescription menu whenever it is reviewed.
b. ACCEPT? YES/NO: Use the up/down arrows to move to a field and change the data
entered at this time. To abort this screen entirely, press “NO” at this prompt, and the
display will return to the options menu. If the rate is correct, press the “YES/ENTER”
key, and the new rate will automatically be accepted. The display returns to the Options
menu when this screen is exited.
c. New values will be in effect when the Options Menu is exited.
5 DN Occlu (Down Occlusion): This feature allows the pump to sense an occlusion pressure
in the line between the pump and the access site— e.g., a kink in the tubing between the
pump and the patient, a blockage in the access site, a clamp left on the distal tubing. There
are two options for this field: the “Low” setting (default setting) triggers the alarm when it