User`s manual
47
The following screen displays the Continuous Therapy “Options” available. When you open this
menu, use the up and down arrows to move the highlight bar to the desired option. Lock settings
may require an access code before the option can be changed. Any new setting, except “CLEAR
HOURLY” screen acceptance becomes effective when the “Options” screen is exited by pressing
“YES/ENTER” at the “ACCEPT OPT?” field.
Continuous Options Menu
Note: The pump’s default settings are indicated above in the left screen.
1 LOCK: When the lock is set to “OFF ” the clinician has access to all of the pump functions.
After the pump is completely programmed, change the lock to meet the security needs of this
therapy. There are four choices: OFF, 1, 2, or 3. Each level higher limits the available
options. To change the lock setting, press the “NO/CHANGE” key until the correct lock
setting appears. Then, press “YES/ENTER” key to accept the new lock selection. Once the
pump has been set into a locked condition, any subsequent request to change the lock will
require a clinician access code. (See “Clinical Information and Access Code” chapter for
access code information, lock setting table and available features for each lock setting).
2 HRLY TOT (Hourly Totals): To view hourly infusion totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. The following screen will appear:
Hourly Totals Screen
The screen displays hourly infusion delivery for each hour (00-23) of the last 24 hours of
therapy in the unit selected.
a. The highlighted field has a marker ◄ indicating the current hour in progress.
b. Use the up arrow key to scroll back through the last 24 hours.
c. If the infusion was not running, a series of ------ will appear for that hour.
Continuous Therapy