User`s manual

36
11 CLEAR Pt. DATA: Selecting this feature clears all data programmed and resets the
pump to its default settings. The history log is NOT cleared. All disabled therapies
are reenabled. When you select “YES/ENTER” at this field, the following screen displays:
When you select “YES,” all patient data clears and the display returns to the BIOMED
SETUP Menu. Selecting “NO” aborts the feature and also returns to the menu.
12 CHANGE CONTRAST: The contrast on the display screen can be adjusted. To do so,
press the “YES/ENTER” key at this field, and the following screen appears:
The level can be adjusted from 0 (lightest screen) to 100 (darkest screen). The pump is set to
an optimized setting when shipped from the factory. To change the setting, enter any
number between 0 and 100, and press the “YES/ENTER” key to see the change in the
screen. If this level of contrast is not satisfactory, move the cursor back to the “LEVEL”
field and enter another number and press the “YES/ENTER” key. When the contrast is
satisfactory, move the cursor to the ACCEPT field, and press the “YES/ENTER” key. To
abort the entire selection and return to the previous contrast setting, press the
“NO/CHANGE” key at this field and return to the BIOMED SETUP Menu.
Changing Date and Time
13 CHANGE DATE AND TIME: When this field is selected, the screen on the left appears:
Change Date and Time Screen
This screen is used to set the system Real Time Clock in the pump and will change the date
and time displayed on the Welcome screen.
Note: The time must be reset to accommodate Daylight Savings Time changes and also at any
time the pump is moved into a different time zone.