to all pump functions. Each level higher limits the options available to the patient or
caregiver (see each lock option screen in the individual therapy sections and/or the lock level
table in the chapter entitled Clinician Information and Access Code.) If the desired lock does not
appear on the display, press the “NO/CHANGE” key until the correct lock does appear.
Then, press the “YES/ENTER” key to accept and move the cursor to the next field.
7 ENABLE THERAPY: This field is used to customize the Curlin Medical pump to the
specific therapy ordered for the patient. When you select this function by pressing the
“YES/ENTER” key, the menu below appears, listing all five of the possible therapies. The
pump’s default setting is to enable all therapies. You can choose, however, to
enable
or
disable
any of the therapies listed. Once a therapy is disabled, no further prompts will appear
in programming for that therapy.
Enable Therapy Menu
a. To enable a therapy, move the cursor to that therapy selection and press the
“YES/ENTER” key when the selection states “ON.”
b. To disable a therapy, move the cursor to that therapy selection and press the
“NO/CHANGE” key. The “ON” status will change to “OFF.” Press the “YES” key to
accept the “OFF” or disable status.
c. When the “ON/OFF” status of each therapy is satisfactory, move the cursor to the last
field, “ACCEPT?” and press the “YES/ENTER” key to accept. To abort the enable
therapy process, press the “NO” key to exit this menu and return to the BIOMED
SETUP Menu.
d. At least one therapy must remain enabled before you exit this screen.
Note: If a therapy that has previously been programmed is disabled, the program is cleared, and
the default settings are reestablished.
Patient History Log
The patient history log is a cumulative record of the approximately last 6,000 events that have
occurred in the pump. Example: When prescription information is saved using the
“YES/ENTER” key at the “DONE” or “ACCEPT” fields, a copy of the information also
gets ‘date and time stamped’ and is recorded in the patient history log. Alarms, error codes,
certain alerts, and certain key press events also get logged when they occur. The patient
history log can be printed using the Curlin Medical print package and by selecting the “Print
Pt. Hx” option. Selecting the “Clear Hx Log” option will erase the patient history log. When
the history log is full, it will overwrite on a “first-in, first-out” basis.
Basics Programming