User`s manual

19 19
External Power Sources
Battery Pack
The battery pack is an external, rechargeable, lead acid battery pack than can be used as an
additional, long-life, portable, external source of power. It is rechargeable by plugging the AC
Adapter into the recharging port. The battery pack is connected to the pump via the two-hole
port located on the bottom of the pump (as shown below). The two-pin connector plugs into the
receptacle correctly by aligning the red dots on each.
Note: Pause the pump and then turn it off before connecting the Battery Pack. Note also that
the connector is keyed with a “notch” at this red-dot location to guide alignment. To remove, pull
back on the locking sleeve to unlock and then pull gently to remove the connector from the
pump’s port.
Figure 1.12 Battery Pack
Introduction
T
o charge the Battery Pack:
Plug the AC Adapter into a grounded wall outlet
and then plug the AC Adapter into the
recharging port of the Battery Pack (see above).
On initial use, the Battery Pack should be
charged for a full 24 hours, subsequent recharge
takes approximately 12 hours (overnight).
Locking Sleeve
Recharging Port
Illustration of the
bottom of the Curlin
Medical Pump
Align the red
dots and the
notches. Then
plug the two-
pin connector
into the two-
hole receptacle
located on the
bottom of the
Curlin Medical
pump.