User`s manual
18 18
Figure 1.11 Installing batteries into the Curlin Medical Ambulatory pump
Whenever an external source is powering the pump, it is still
important to always install the two “C” size batteries. In the event of a
power failure or accidental power cord disconnection, the batteries
will power the pump without interrupting the therapy in progress. If
the batteries are in place when the power cord is disconnected, the
display will say “External Power Disconnected,” the alert beep will
sound, the batteries will take over powering the pump, and the
therapy will continue infusing.
Two “C size”
alkaline
batteries
The battery compartment door is located
on the back of the Curlin Medical pump.
To open the door, depress the recessed
release pin (it may be helpful to use a
small coin to depress the pin), slide the
battery door down, and remove it.
Replace the old batteries with two new
“C” size alkaline batteries. Install them
by depressing the coil spring with the
bottom of the battery and sliding the top
in place. Note: both positive poles are at
the top as shown in Figure 1.11. Always
replace both batteries. Slide the battery
compartment door back to its closed
position until the recessed pin locks into
place.
PLEASE NOTE:
Follow local governing ordinances and
recycling plans regarding disposal or
recycling of device components. When
the “C” size batteries are discarded do
not incinerate them.
Battery
Compartment
Door
Recessed Release
Pin
Coil Springs
IMPORTANT
INFORMATION