Operating Time with Full Charge
Batteries:
Nominal 10 hours at a rate of 400 ml/hr
Nominal 30 hours at a rate of 125 ml/hr
Nominal 85 hours at a rate of 2 ml/hr
Internal Memory Battery 5 years life
Display Graphic LCD, 4 text fields, 16 characters per
field
Flow Rates 0.1 to 400 ml/hr
KVO Rates 0 to 10 ml/hr. Can be set at “0” in Continuous,
TPN, Variable and Intermittent Therapies
Volume Accuracy
± 6% for volume delivery >1 ml under the
following conditions:
• Temperature 55° to 90° F
• Back pressure to 300 mmHg (5.8 PSI)
±10% for volume delivery >1 ml under the
following conditions:
• Temperature 45° to 55°F
• Temperature 90° to 104°F
• Back pressure to 750 mmHg (14.5 PSI)
Recommended administration set usage not to
exceed 72* hours.
*follow standards of practice for set changes
Administration Sets Proprietary Curlin Medical non-DEHP, PVC
administration sets available in different
configurations.
Fill Volume for Administration Sets Nonfiltered microbore sets ± 4 ml.
See individual administration set label for specific
fill volume
Up Occlusion Detection Time
Microbore tubing with occlusion
below spike
Rate
0.1 ml/hr
1 ml/hr
10 ml/hr
50 ml/hr
Time to Alarm
<20 minutes
< 4 minutes
<20 seconds
<10 seconds
Down Occlusion Detection Time
Microbore tubing with occlusion at
distal Luer
Rate
0.1 ml/hr
1 ml/hr
10 ml/hr
50 ml/hr
125 ml/hr
Low Pressure
<45 min
<5 min
<20 seconds
< 5 seconds
< 3 seconds
High Pressure
<1 hr, 30 min
<7 min
<45 seconds
<15 seconds
< 5 seconds