User`s manual

149
d Program the following CONTINUOUS Infusion:
Pre Rx:
UNITS: ml
DELAY: OFF
Titrate: OFF
NEXT? YES
Prescription:
BAG VOL: 100 ml
VOL TBI: 10.0 ml
RATE: 125 ml/hr
TIME: 0:04 HH:MM
KVO Rate: 0.0 ml/hr
DONE? YES
e Start the continuous therapy running by pressing RUN button.
f When the 10 ml infusion is over, observe the amount volume infused as measured
on the measurement device. Passing measurement is: 9.5 g to 10.50 g (9.5ml to
10.5ml).
g When Infusion complete occurs verify that the Audio Alarm is beeping and that the
Green and Red LEDs are flashing.
3 CHECK ALL SAFETY SYSTEMS:
a Low Down Stream Occlusion - Repeat the same Continuous Infusion as above.
Go to the Option screen and confirm that the Down Occlusion level is set to LOW.
Start the infusion and while the pump is running pinch off or close the down stream
side of the administration set. Confirm the pump goes into Down Occlusion alarm.
Open or clear the down stream occlusion and confirm the pump resumes infusion.
b High Down Stream Occlusion – Go to the Option screen and change the Down
Occlusion level to HIGH. Again pinch off or close the clamp on the down-stream
side of the administration set. Confirm the pump goes into Down Occlusion. Open
or clear the down stream occlusion and confirm the pump resumes infusion.
c Up Stream Occlusion - Resume the infusion pumping then pinch off or close the
up-stream side of the administration set. Confirm that the pump goes into an Up
Occlusion alarm. Open or clear the up stream occlusion, press Pause then Resume
the infusion.
d Door Sensor - With infusion in progress open the door latch and observe ALARM
DOOR OPEN OR SET NOT INSTALLED.
e Air In Line Sensor - Reinstall the administration set properly and resume the
infusion. From the Options menu set the AIL SENS (AIR IN LINE sensitivity) to
0.1ml. With the pump running introduce an Air Bolus of approximately 0.5ml into
the fluid path at the input side of the pump. Once approximately 0.1ml of air has
past the AIL detector, an ALARM AIR IN LINE will occur. Pause and Prime the
air bolus past the AIL sensor, then resume the infusion.
f Auxiliary Alarm - With the pump running observe the Vol INF (volume infused on
the run screen) and disconnect the pump from all power sources (unplug the AC
Cleaning, Maintenance