User`s manual
146
Transport and Storage
During transport or storage, the batteries should be removed, the pump should be cleaned and
kept in its hard-shell packing case. The environmental conditions should be within the range of:
• An ambient temperature -20° F to 140°F
• A relative humidity range of 10% to 95%
• An atmospheric pressure range of 500 Hpa to 1060 Hpa
All accessories should be stored in their original packing containers to ensure the longest service.
Operating Environment
The following conditions of ambient temperature range and relative humidity are recommended
for the fluid delivered accuracy of:
• Temperature: ±6% Accuracy: 55 to 90°F
±10% Accuracy: 45 to 55°F
±10% Accuracy: 90 to 104°F
• Relative Humidity: 15 to 95% Non-condensing.
Maintenance
Yearly preventive maintenance checks are recommended and will be prompted in the “Set-Up”
menu and on the display when the date occurs. This date accrues from the date of factory release
and is reset when each annual maintenance check is completed.
Any time an error code message occurs repeatedly or any time a pump appears to be tampered
with or shows any indication of not performing to standards, the pump should be returned to the
healthcare provider dispensing the pump.
At no time should unauthorized personnel open the hard case of the pump.
Limited Warranty
Curlin Medical, LLC (the “Manufacturer”) warrants to the Original Purchaser that
the infusion pump (the “Pump”), not including accessories, shall be free from
defects in material and workmanship under normal use, if used in accordance with
this User’s Manual, for a period of one year from the actual date of sale to the
Original Purchaser. THERE ARE NO OTHER WARRANTIES.
The warranty does not cover normal wear and tear and maintenance items, and specifically
excludes batteries, administration sets, extension sets or any other accessory items or
equipment used with the Pump.
Subject to the conditions of and upon compliance with this Limited Warranty, the
Manufacturer will repair or replace at its option without charge (except for a minimal charge
for postage and handling) any Pump (not including accessories) which is defective if a claim
is made during such one year period.
The following conditions, procedures, and limitations apply to the Manufacturer’s obligation
under this warranty: