User`s manual
140
Battery Pack (340-2006)
This accessory is an external, rechargeable, lead-acid battery pack than can be used as an
additional, long-life, portable, external source of power. It is recharged with the AC Adapter.
(Plug the AC Adapter into a wall receptacle and then into the recharging port of the battery pack).
The battery pack is connected to the pump via the two-hole port located on the bottom of the
pump (as shown below). The two-pin connector plugs into the receptacle correctly when the red
dots are aligned on each. Note, also, that the connector is keyed with a “notch” at this red-dot
location to guide alignment. To remove, pull back on the locking sleeve to unlock, and then
remove the cord from the pump’s port. When using this power source, the power status screen
will indicate “External Power Source”.
Figure 10.2 Battery Pack
Soft Carry Packs
Curlin Medical has some recommended carry packs that come in four sizes and are
designed to provide user comfort and convenience. Please refer to our website at
www.curlinmedical.com to review the latest of these carry packs.
T
o charge the Battery Pack:
Plug the AC Adapter into a grounded wall outlet
and then plug the AC Adapter into the
recharging port of the Battery Pack (see above).
On initial use, the Battery Pack should be
charged for a full 24 hours, subsequent recharge
takes approximately 12 hours (overnight).
Locking Sleeve
Recharging Port
Illustration of the
bottom of the Curlin
Medical Pump
Align the red
dots and the
notches. Then
plug the two-
pin connector
into the two-
hole receptacle
located on the
bottom of the
Curlin Medical
pump.