User`s manual
135
Error
Code
LCD Displayed
Message
Probable Cause Corrective Action
34 Turn Pump OFF
Call Provider
Audio Auxiliary Speaker High Report error code to factory.
35 History File
Turn pump OFF/ON
History Log Erased Report error code to factory.
36 Bad Cal Data
Turn pump OFF
Call Provider
Bad EEPROM Data CRC
value
Report error code to factory.
37 Turn Pump OFF
Call Provider
Analog to Digital Converter did
not Run
Report error code to factory.
38 Turn Pump OFF
Call Provider
Air In Line Circuit not Tested Report error code to factory.
39 Turn Pump OFF
Call Provider
Bad EEPROM Data CRC 2 Report error code to factory.
40 Turn Pump OFF
Call Provider
Bad EEPROM Data Header Report error code to factory.
41 Bad Cal Range
Turn Pump OFF
Call Provider
Bad EEPROM Limit Report error code to factory.
100 Turn pumpOFF/ON
Call Provider
Menu not found Report error code to factory.
101 Turn pumpOFF/ON
Call Provider
Mode not found Report error code to factory.
102 Turn pumpOFF/ON
Call Provider
Item not found Report error code to factory.
103 Turn pumpOFF/ON
Call Provider
Field not found Report error code to factory.
104 Turn pumpOFF/ON
Call Provider
Data mismatch Report error code to factory.
105 Turn pumpOFF/ON
Call Provider
No menu pointer Report error code to factory.
106 Turn pumpOFF/ON
Call Provider
Count out of range Report error code to factory.
107 Turn pumpOFF/ON
Call Provider
Display task queue overflow Report error code to factory.
108 Turn pumpOFF/ON
Call Provider
Divide by zero Report error code to factory.
109 Turn pumpOFF/ON
Call Provider
Concentration zero value Report error code to factory.
110 Turn pumpOFF/ON
Call Provider
Invalid occurrence Report error code to factory.
111 Turn pumpOFF/ON
Call Provider
Multiple param struct access Report error code to factory.
112 Turn pumpOFF/ON
Call Provider
Multiple status struct access Report error code to factory.
140 Turn pumpOFF/ON
Call Provider
Infusion volume out of range Report error code to factory.
141 Turn pumpOFF/ON
Call Provider
Infusion rate out of range Report error code to factory.
142 Turn pumpOFF/ON
Call Provider
Infusion time out of range Report error code to factory.
143 Turn pumpOFF/ON
Call Provider
Volume conversion error Report error code to factory.
144 Turn pumpOFF/ON
Call Provider
Rate conversion error Report error code to factory.
145 Turn pumpOFF/ON
Call Provider
Time conversion error Report error code to factory.
146 Turn pumpOFF/ON
Call Provider
Bad PCA dose ratio Report error code to factory.
160 Stuck PERIOD Key
Turn pump OFF/ON or Call
Provider
Stuck Period / Silence Key If key was pressed, release it.
Otherwise report error code to
factory.
161 Stuck ON OFF Key
Turn pump OFF/ON or Call
Provider
Stuck On / Off Key If key was pressed, release it.
Otherwise report error code to
factory.
Troubleshooting