User`s manual
133
LCD Display: UPLOAD CkSum ERR
Alarm: CMS Upload experienced a Check Sum Error
Cause: At the completion of an upload of a CMS script the Check Sum signature
from the CMS system did not match the signature calculated by the pump
Resolution: A second attempt to upload the script should be made. If the second
attempt fails, then a new script should be created for the desired
prescription. (See CMS User’s Manual).
Troubleshooting