User`s manual

127
Resolution: Press “PAUSE” key before pressing “OFF” key to stop an infusion.
LCD Display: ALERT Rx IN PROGRESS, YES to Repeat – NO to Return
Alert: REPEAT Rx selected while a therapy is already in progress
Cause: REPEAT Rx was selected and there is a therapy currently in progress.
Resolution: To terminate the current therapy in progress, press YES. Resume will no
longer be available. The Select Therapy menu will be displayed (if only
one therapy is in programmed, the pre Rx menu for the therapy will be
displayed). When the therapy to repeat is chosen, the last prescription data
programmed will appear in the menu, excluding Delayed Start and Load
Dose (if therapy was previously started). When the infusion is started the
entire therapy will be repeated.
To continue the current therapy in progress, press NO. The Pause menu
will be displayed.
LCD Display: ALERT Rx IN PROGRESS, YES New Program – NO to Return
Alert: NEW PROGRAM was selected while a therapy is already in progress
Cause: NEW PROGRAM was selected and there is a therapy currently in
progress.
Resolution: To terminate the current therapy in progress, press YES. Resume will no
longer be available. The Select Therapy menu will be displayed (if only
one therapy is in programmed, an alert screen informing that the Rx will
be erased will be displayed). When the therapy for the new program is
chosen, entry of a new prescription will be required.
To continue the current therapy in progress, press NO. The Pause menu
will be displayed.
LCD Display: ALERT Rx Programmed, YES to Erase – NO to Return
Alert: NEW PROGRAM was selected for a therapy that has an existing
prescription
Cause: NEW PROGRAM was selected for a therapy that has an existing
prescription.
Resolution: To erase the prescription for the selected therapy, press YES. The pre Rx
menu will be displayed and entry of a new prescription will be required.
To preserve the prescription for the selected therapy, press NO. The
Pause menu will be displayed.
LCD Display: ALERT SET NOT INSTALLED
Alert: Set Not Installed
Cause: Administration set is not properly installed.
Resolution: Install the appropriate Curlin Medical administration set for the desired
therapy. Use ONLY Curlin Medical administration sets. See Chapter 1,
Troubleshooting