Resolution: Wait until the dose in progress completes before modifying field.
LCD Display: ALERT NOT in PCA. NO BOLUS ALLOWED
Alert: Bolus requested when therapy is not PCA
Cause: Remote Bolus button or BOLUS key is pressed when not in PCA therapy.
Resolution: Remove the remote bolus cord from the pump if connected. Refrain from
pressing the BOLUS key.
LCD Display: ALERT NUMBER OF DOSES INVALID
Alert: Number of doses entered exceeds maximum for therapy
Cause: The number of doses per bag in Intermittent therapy exceeds the allowable
maximum.
Resolution: Recheck data entry. The number of doses requested must be less than or
equal to the maximum for the therapy.
LCD Display: ALERT PERFORM ROUTINE MAINTENANCE
Alert: Maintenance Date (displayed only if date has arrived when pump is turned
on)
Cause: Current Date is beyond the Maintenance Date.
Resolution: Return pump for routine maintenance.
LCD Display: ALERT RAMP INVALID FOR AMT/RATE
Alert: Ramp time entered is invalid for amt/rate already programmed
Cause: Ramp value entered for TPN is unacceptable given the desired rate and
amount programmed.
Resolution: Recheck all data entered for prescription. If programming a new TPN
therapy (”NEW PROGRAM” selected) and the ramp time is a
requirement, enter the ramp time(s) and total time first followed by the
next governing parameter (rate or volume to be infused).
LCD Display: ALERT RATE EXCEEDS DELTA RATE/DELTA TIME
Alert: Rate entered exceeds the delta time limits
Cause: The rate entered falls outside of the allowable range for the delta rate /
delta time limits programmed.
Resolution: Recheck data entry or wait more time elapses then attempt to change the
rate. Rate selected cannot exceed the range placed on titration.
LCD Display: ALERT RUNNING, PAUSE PUMP BEFORE OFF
Alert: Pause Infusion before turning off
Cause: The OFF key was pressed while the pump was infusing or a running
screen displayed.