User`s manual

125
LCD Display: ALERT NEW DOSE IS LESS THAN CURRENT DOSE
Alert: Unable to program a lower number of doses than number of doses already
infused
Cause: While the Variable or Intermittent therapy is in progress, a lower number
of doses to be delivered is entered than the number already infused.
Resolution: Wait until current therapy in progress is complete, or “PAUSE” pump
and, using the REPEAT Rx function, modify the number of doses.
Note: Selecting to repeat the therapy in progress will terminate the current
therapy.
LCD Display: ALERT NO BOLUS DOSE DEFINED
Alert: Unable to enter bolus interval or # of bolus/hr because no bolus value has
been programmed
Cause: Entry of a value other than zero for the Bolus Interval or #Bolus per hour
field when no bolus value has been programmed.
Resolution: Enter bolus value before entering bolus interval or # of boluses/hr.
LCD Display: ALERT NO KEYPAD ACTIVITY
Alert: No Keypad Activity
Cause: The pump has been displaying a non-running screen for two minutes
without any keys being pressed. The alert will recur every 2 minutes as
long as there are no key presses.
Resolution: Complete the activity for the displayed menu. The pump may also be
turned off, however, all programming that was not complete will be
discarded.
If the pump was running and the Options menu displayed when the alert
occurred, the run screen for the therapy will be displayed when the alert is
cleared. Any changes to the Options menu are discarded and will have to
be re-entered.
LCD Display: ALERT NO THERAPY ENABLED
Alert: No Therapy Enabled
Cause: Attempting to accept the Enable Therapy menu with all therapies set to
“OFF”. At least one therapy must be enabled or “ON” for menu to be
accepted.
Resolution: Enable the therapy or therapies that are to be active for the pump.
LCD Display: ALERT NOT ALLOWED, DOSE IN PROGRESS
Alert: Change is not allowed while a dose is in progress
Cause: Modification to the current field is not allowed while the dose is in
progress.
Troubleshooting