LCD Display: ALERT KVO RATE IS GREATER THAN INF RATE
Alert: Unable to enter KVO rate that is greater than infusion rate
Cause: Occurs when user enters a KVO rate that is greater than infusion rate.
Resolution: Recheck data entry. KVO rate must be less than programmed or
calculated infusion rate.
LCD Display: ALERT LESS THAN MIN FIELD LIMIT, MIN = ________
Alert: Unable to program a field with a value that is less than the minimum limit
Cause: Value entered is less than the minimum limit for the pump or titration
limit programmed.
Resolution: Recheck data entry. A value entered cannot be less than the field’s
minimum limit.
LCD Display: ALERT “LOW C BATTERIES” or “LOW EXTERNAL BATTERY”
Alert: Battery source power is low
Cause: The batteries that are powering the pump have reached levels that may
prevent the continued operation of the pump beyond 1 hour (actual time is
rate dependent). Alert applies to both “C size” and rechargeable battery
pack.
Resolution: “PAUSE” the therapy in progress, turn the pump “OFF”; and then replace
the batteries. (See chapter 1 of the User’s Manual for instructions on
battery change.) Replace or recharge the battery pack if applicable.
LCD Display: ALERT MAX CHANGES FOR PERIOD REACHED
Alert: The maximum number of changes for the current delta time window has
been reached
Cause: The maximum number of pump rate changes for the current delta time
window has been reached. This occurs when the rate has been changed
more than 12 times during any delta time window.
Resolution: Allow more time to elapse before attempting to change the rate or
reprogram the therapy with titration limits and prescription basal and bolus
ranges that will reduce the number of times that the rate will have to be
adjusted.
LCD Display: ALERT MODIFICATION NOT ALLOWED
Alert: Modification Not Allowed
Cause: The selection cannot be modified at this time or the programmed lock
setting does not permit modifications to this selection.
Resolution: Review the User’s Manual for program restrictions or if lock level
restricted, change lock setting to access area you wish to modify.