User`s manual

123
Resolution: Recheck the field and accept it by pressing the “YES/ENTER” key.
LCD Display: ALERT GREATER THAN MAX FIELD LIMIT, MAX = _____
Alert: Unable to program a field with a value that exceeds the maximum limit
Cause: Value entered is greater than the allowable limit.
Resolution: Recheck data entry. A value entered cannot exceed the field limit.
LCD Display: ALERT GREATER THAN WHAT BOLS INT ALLOWS
Alert: Unable to enter a # of boluses/hr that exceeds the possible number
allowed given the bolus interval
Cause: # Boluses/hr value exceeds the possible number given the bolus interval
(“BOLS INT”) programmed.
Resolution: Recheck data entry. The # of boluses/hr cannot be set for higher than the
number allowed using the given bolus interval.
LCD Display: ALERT INFUSION NOT IN PROGRESS
Alert: Infusion not in progress
Cause: Modification attempted to Next Dose or Early Down Ramp field without
an infusion in progress.
Resolution: Wait until infusion is in progress before modifying or modify original Rx
programmed.
LCD Display: ALERT INSUFFICIENT VOLUME MAX DOSES=_____
Alert: Bag Volume value is insufficient to deliver the remaining
Cause: Occurs whenever the prescription is changed so that Bag Volume is now
less than Amount Required for the entire therapy (based on volume
already infused plus volume required to deliver the remaining doses).
Resolution: Recheck data entry. If the source container contains sufficient volume to
complete the remaining dose cycles (dose volume + KVO volume),
increase the “Bag Vol”. If the volume is insufficient, decrease the
“DOSE/BAG” such that the therapy can continue with the new modified
prescription.
Note: If either “Bag Vol” or “DOSE/BAG” is modified and the therapy
is repeated, these fields will have to be verified against the source
container volume and prescribed prescription and changed
accordingly.
LCD Display: ALERT INVALID ENTRY
Alert: Invalid Entry
Cause: Data entered does not fall within the defined limits of the program.
Resolution: Enter data that falls within the limits of the program. Refer to the specific
therapy section in the User’s Manual.
Troubleshooting