User`s manual
121
If it is not necessary to preserve the therapy in progress or any other
programmed therapy, use the “CLEAR Rx” feature of the Clinician
Biomed menu. This feature will erase all programmed therapies in the
pump. The Date and Time can then be changed.
If the therapy is to be terminated but the prescription preserved (excluding
Delayed Start), go to the “Pause” menu, select “REPEAT Rx” and press
“YES” at the “Rx in progress” alert screen. The pump can then be turned
off and then on to access the Date/Time change option in the Clinician
Biomed menu. Selecting “NEW PROGRAM” and the name for the
therapy in progress, instead of “REPEAT Rx”, may also be used but this
selection will erase the prescription for the therapy.
LCD Display: ALERT DATE/TIME INVALID
Alert: Unable to enter invalid date or time
Cause: Date or time entry is not in a valid format.
Resolution: Recheck data entry. An invalid date or time will not be accepted.
LCD Display: ALERT DELAY START TIME NOT VALID
Alert: Delay Start time not valid
Cause: Delay start time entered is less than one minute or greater than 24 hours
from the current system time.
Resolution: Reset Delay Start Time to fall within pump limits. (To obtain the current
date and time of the pump, turn the pump off and then on and observe
the time on the Welcome screen).
LCD Display: ALERT DOOR OPEN or SET NOT PROPERLY INSTALLED
Alert: Door Open or Administration Set is not properly installed
Cause: Door Latch is not completely closed or the Administration Set is not
installed according to installation instructions.
Resolution: Recheck the door latch assembly and securely close the door. If the door
was closed properly, re-install the set according to the installation
instructions.
LCD Display: ALERT DOSE AMT IS LESS THAN DOSE INFUSED
Alert: Dose amount entered is less than the amount already infused for the dose.
Cause: New dose value entered is less than the amount already infused for the
dose.
Resolution: Recheck data entry. After the dose in progress finishes, the dose amount
may be changed to the lower value.
If the above is not suitable, re-program the prescription.
LCD Display: ALERT DOSE CANCELLED, RESUME INFUSION, THEN NEW
DOSE
Troubleshooting