User`s manual

120
LCD Display: ALERT CALCULATED RATE OUT OF RANGE
Alert: Invalid calculated rate of infusion
Cause: The calculated rate for the infusion, based on the values entered for
Volume/Amount and Time, exceeds either the minimum or maximum
rate limit of the pump or the maximum rate programmed in titration
limits.
Resolution: Recheck data entry for Rx.
LCD Display: ALERT CALCULATED TIME OUT OF RANGE
Alert: Invalid calculated time
Cause: The calculated time for the therapy, based on the values entered for
Volume/Amount and Rate, exceeds the therapy time limit of 288:00
HHMM (167:59 HHMM Software Versions below R37C).
Resolution: Recheck data entry. The time calculated by the pump cannot exceed the
pump range.
LCD Display: ALERT CHANGES REQUIRE REPROGRAMMING. YES to Continue
NO to Return
Alert: Changes to the data will require re-entry of prescription.
Cause: Attempted to change the Pre-Prescription Screen prior to start of infusion.
Resolution: If a change to the programmed prescription is required, press “YES”.
Note: Selecting “YES” will clear the entire
prescription and require re-
entry the prescription.
If no change is required, press “NO” to preserve the current prescription
and return back to the Pre-Prescription screen.
LCD Display: ALERT CMS LEARN MODE ACTIVE – KEY NOT ACCEPTED
Alert: The pump is performing a CMS Create Prescription function and the key
pressed is not valid for this mode.
Cause: While using the pump with CMS/PC in order to create a prescription only
certain keys are allowed to be recorded as part of the prescription.
Whenever a prohibited key is pressed (i.e., RUN key) this alert will occur
and the key will not be recorded.
Resolution: Do not press the key causing the alert during a CMS Create Prescription
function.
LCD Display: ALERT DATE/TIME CHANGE NOT ALLOWED
Alert: Not allowed to change date or time
Cause: Attempt made to modify system clock while a therapy/delay is in progress.
Resolution: If the therapy in progress is to be continued, wait until the infusion is
complete and then change the Date and Time.