If no air-eliminating filter is being used, press NO. The infusion will
start/resume and the Air In Line sensitivity will be set to 2 ml for the
remainder of the infusion.
Note: If the screen is not acknowledged with a YES or NO key press, the
Air In Line sensitivity will be set to 2 ml when the screen is cleared
(i.e. alarm occurs while screen is displayed).
LCD Display: ALERT AIR SENSOR HAS BEEN SET TO 2.0 ML
Alert: Air in line sensitivity will be set to 2.0ml
Cause: NO was pressed at the “AIR SENS OFF. Using In Line Filter?” alert
screen. The air in line sensitivity will be set to trigger an AIR IN LINE
alarm if 2.0ml of air is detected.
Resolution: If an air eliminating filter is being used, press YES when the “AIR SENS
OFF. Using In Line Filter?” screen is displayed. This will disable air
detection as indicated under “AIR SENS OFF. Using In Line Filter?” alert
explanation.
LCD Display: ALERT AMT NOT 0 OR CURRENT DOSE AMT
Alert: Unable to change clinician dose while clinician dose in progress
Cause: A clinician dose is in progress, and a new clinician dose value is entered
that is not “0” or the current dose amount.
Resolution: If the current clinician dose is to be continued, change the ‘Clin DOSE’
value back to the original value and accept the field and menu.
If the current clinician dose is to be terminated, which will allow a new
clinician dose to be programmed, enter a value of 0 in the ‘Clin DOSE’
field and accept the field and menu. The next time that the Options menu
is accessed a new ‘Clin DOSE’ value can be entered.
LCD Display: ALERT AMT TBI GREATER THAN BAG VOLUME
Alert: Value entered is greater than value in Bag Vol field.
Cause: Occurs when the amount the entered for the “Amt TBI” or “Vol TBI”
field is greater than the value in the bag volume field.
Resolution: Recheck the data entry.
LCD Display: ALERT AMT TBI GREATER THAN BAG VOLUME. YES to
Continue NO to Return
Alert: Insufficient volume
Cause: Occurs when bag volume is less than amount to be infused. Can occur
when pump’s priming feature is used as bag volume is depleted by the
amount used in priming.
Resolution: To modify the Bag Volume or Amt/Vol TBI, press “NO” to return to the
Run Options screen. The programming screens will then be accessible by
pressing “NO” to Review on the Run Options screen.