User`s manual
113
stop the auto-advance function by pressing any key (except ON/OFF) and then complete the
confirmation of the prescription manually by pressing the “YES” key at the remaining fields.
When the program is accepted, the display returns to the above “Run Options Screen.” If the
current therapy is still in progress when the “REPEAT Rx” function is selected, the following
screen will appear:
a. This screen is a reminder that the infusion is not complete. If the therapy is to be
terminated and a new IV container added, press “YES/ENTER” to go to the
programming screen, confirm the program and proceed with the repeat infusion.
b. If the incomplete therapy is to be continued, press “NO” and the Pause Menu screen will
be displayed, then press “YES” and the Run Options screen will be displayed. To resume
the unfinished therapy, press “RUN”.
Entering a New Program
3 NEW PROGRAM: To program an entirely new prescription or different therapy, press the
“YES/ENTER” key at this field (The ability to change these parameters may be controlled
by the lock setting). The display will return to the “Select Therapy Screen” if more than one
therapy mode is enabled, or to the “Pre-Prescription Screen” of the single therapy mode. If
the current container has not completely infused, the following screen will appear:
a. This screen is a reminder that the infusion is not complete. If you want to return to the
current container and finish infusing it, press the “NO/CHANGE” key to return to the
Pause Menu screen.
b. To abort the current container and enter a new program, press the “YES/ENTER” key.
If a prescription has previously been entered for the therapy mode you wish to program,
the following screen will appear:
c. If you want to save the current prescription, press the “NO/CHANGE” key to return to
the Pause Menu screen. If you wish to erase the current prescription and enter a new
program, press the “YES/ENTER” key to proceed. All previous prescription values will
be erased.
Variable Therapy