User`s manual

108
1 LOCK: When the lock is set to “OFF, ” the clinician has access to all of the pump functions.
After the pump is completely programmed, change the lock to meet the security needs of this
therapy. There are four choices: OFF, 1, 2, or 3. Each level higher limits the available
options. To change the lock setting, press the “NO/CHANGE” key until the correct lock
setting appears. Then, press “YES/ENTER” key to accept the new lock selection. Once the
pump has been set into a locked condition, any subsequent request to change the lock will
require a clinician access code. (See “Clinical Information and Access Code” chapter for
access code information, lock setting table and available features for each lock setting).
2 HRLY TOT (Hourly Totals): To view hourly infusion totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. The following screen will appear:
Hourly Totals Screen
The screen displays hourly infusion delivery for each hour (00-23) of the last 24 hours of
therapy in the unit selected.
a. The highlighted field has a marker indicating the current hour in progress.
b. Use the up arrow key to scroll back through the last 24 hours.
c. If the infusion was not running, a series of ------ will appear for that hour.
d. To exit this screen, press the NO/CHANGE key.
3 CLR HRLY (Clear Hourly Totals): To clear Hourly Totals press the “NO/CHANGE” key
to change OFF to ON and then press “YES/ENTER”. The following screen will appear:
Clear Hourly Totals Screen
a. To clear the Hourly Totals, press the “YES/ENTER” key. All Hourly Totals will be
cleared and the cursor will return to the Options menu.
b. To maintain the Hourly Totals, press the “NO/CHANGE” key. Hourly Totals will be
maintained and the cursor will return to the Options menu.
c. A lock level of 2 or 3 will require an access code entry before allowing “Hourly Totals”
reset.