User`s manual

101
Variable Therapy
This therapy is designed to allow twenty-four specified programs of varying amounts, rates, and times of delivery.
egin programming a Variable Therapy by selecting “Variable” from the “Select Therapy
Screen”. The following display screen will appear, requiring you to enter information
needed to program this therapy.
Prescription Menus
Variable Pre-Prescription Menu Screen
These menu items allow you to customize the program to match each individual prescription.
The pump’s default settings are displayed in the left screen. Each field is individually explained on
following pages.
Note: Some of the lines are abbreviated to fit on the display screen; the full descriptions appear
below.
1 UNITS: Allows you to select from units of ml, mg, or mcg, to match the prescription. The
pump will allow a volume range of 0.1 to 400 ml/hr; therefore, if mg or mcg is selected, the
concentration requested in the next field will determine the upper and lower programming
limits for mg and mcg. Use the “NO/CHANGE” key until the desired “units” appears on
the display, and then press the “YES/ ENTER” key to accept and move to the next field.
2 CONCEN (Concentration): This field is active whenever mg or mcg is selected in the
“units” field and requires data entry of the concentration of mg or mcg per ml. When the
correct data is entered using the numeric keys, press “YES/ENTER” to accept, and move
the cursor to the next field. If ml is chosen in the “units” field, this line will not display. The
range for this field is 0.1 to 999.
3 DELAY Start: This field allows you to delay the start of the infusion from one minute up to
twenty-four hours. If the infusion is NOT to be delayed, select “OFF” and press
“YES/ENTER” to enter. The cursor will move to the next field. If the infusion IS to be
delayed until a later time, press the “NO/CHANGE” key. The delay start will change to
CHAPTER
8
B
Variable Therapy