[ForcePlate_1] (and [ForcePlate_2])
Type Force Plate type: 0=User defined, 1=9281B Kistler, 2=9286 Kistler…, 10=AMTI
BP2416, 11=AMTI-LG6, 12=AMTI-OR6-7, 20=AMTI AccuSway/AccuGait, etc.
Rotation The orientation of the plate relative to Coda. May be 0, 90, 180, or 270 degrees
relative to normal (Force-plate x-axis = Coda Y-axis (perpendicular horizontal),
y-axis = X-axis (parallel horizontal), x-axis = Z-axis (vertical)).
MatThickness The thickness of any covering attached to the surface of the force plate.
CentreOffset Distance (x,y,z) in mm from the Coda marker origin.
ParameterFile Force plate calibration parameters. (Centre of pressure correction coefficients
for Kistler force plates; Calibration matrix for AMTI force plates).
CalFile ADC channel calibration factors (Force plate gains and offsets).
DSPboard The ADC-64 acquisition unit or cPCI A/D board to which the force plate is
connected together with the channel start number (A1:1=ADC_1:Channel start
1…, etc).
BridgeVoltage The AMTI/Bertec bridge excitation voltage (1.0 – 15.0).
AmpGain The AMTI/Bertec amplifier gain (low, high).
CommPort The serial communications port being used on the host computer, laptop or
active hub (1=COM1, 2=COM2…, etc.) for portable force plate connection.
[EMG_1]
Type EMG system type: 3=Noraxon Telemyo 2400T.
Transmitters Specifies the maximum number of transmitters in use.