User's Guide

Cochlear Limited 201136 323649 Iss 3 CP900 Series Sound Processors | User guides
Charge
CP900 Series Sound Processors | User guides 37Cochlear Limited 2011
Lock the CP920 processing unit
To lock, turn the locking screw
clockwise with the CP800
screwdriver until it is in a horizontal
position.
To unlock, turn the locking screw
anticlockwise until it is in a
vertical position.
1
2
Lock and unlock the tamper resistant
battery cover
The standard battery has a tamper resistant lock to help prevent children
opening the battery cover.
Push the lock to the far left with
your Cochlear screwdriver to
lock the battery module.
Childproofing Tip
Push the lock to the far right to
unlock the battery module.
1
2
Angle and fi t the rechargeable
battery into a socket on the
Cochlear Nucleus Battery Charger.
Twist clockwise to connect.
Connect the battery charger to the
power adaptor, plug it into mains
power and switch on.
1
2
3
Charge rechargeable battery modules
Battery wall power light What it means
Steady green
Battery charger power is on
Does not light up Power is not plugged in or turned on at the
power socket, or there is no power.
Battery module light What it means
Steady orange
Battery module is recharging.
Steady green
Battery module is fully recharged.
Flashing orange
There is a problem with the battery module.
Does not light up Battery module is not properly placed, or
there is no power.
This image will show lights.