Troubleshooting guide

Troubleshooting Kit
30
Presented Complaints:
No sound/intermittent sound (help messages H1, H2, H3, H4)
Check that the batteries are correctly inserted.
Turn the sound processor off and then on.
Change the batteries (ensure that all of the batteries are changed).
Check using the lapel microphone.
Check that the coil is fully inserted in the processing unit socket
(and correctly located above the Implant).
Change the coil.
Change the bodyworn controller (can also try the BTE controller).
Change the processing unit (use back-up processing unit or contact the
child’s cochlear implant clinician).
Sound unclear/muffled
Check the LCD for help message (H4 = see steps 2 – 5,
H5 = contact your clinician).
Check using the lapel microphone. Determine whether
the sound improves.
Disconnect the lapel microphone and remove the microphone cover.
Determine whether the sound improves.
Replace the microphone cover, if the sound does not improve.
Replace the processing unit (use back-up processing unit or contact the
child’s cochlear implant clinician).
LCD blank (but sound OK)
Change the bodyworn controller (return the suspected faulty controller).
Sounds uncomfortably loud
Contact the child’s cochlear implant audiologist.
Indicator light
When the indicator light is turned on, it:
Flickers when incoming sounds are being received.
Shows a slow flash rate as a low battery warning.
Shows a fast flash rate for all other warnings e.g. the coil is off its position
over the implant, or there is a sound error (help message H4).
To stop the alarm, push any button.